Instrument: CTD Sea-Bird SBE 911plus

Instrument Short Name: CTD SBE 911plus
(http://vocab.nerc.ac.uk/collection/L22/current/TOOL0058/)
Instrument Description:

The Sea-Bird SBE 911 plus is a type of CTD instrument package for continuous measurement of conductivity, temperature and pressure. The SBE 911 plus includes the SBE 9plus Underwater Unit and the SBE 11plus Deck Unit (for real-time readout using conductive wire) for deployment from a vessel. The combination of the SBE 9 plus and SBE 11 plus is called a SBE 911 plus. The SBE 9 plus uses Sea-Bird's standard modular temperature and conductivity sensors (SBE 3 plus and SBE 4). The SBE 9 plus CTD can be configured with up to eight auxiliary sensors to measure other parameters including dissolved oxygen, pH, turbidity, fluorescence, light (PAR), light transmission, etc.). more information from Sea-Bird Electronics

PI supplied instrument name: CTD Seabird 911plus
Dataset-specific description

Instrument Configuration File for SO-GasEx:

PSA file: C:GasExacqSeasave.psa
Date: 02/23/2008
Instrument configuration file: C:GasExconssec_24_1.con
Configuration report for SBE 911plus/917plus CTD
------------------------------------------------
Frequency channels suppressed : 0
Voltage words suppressed      : 0
Computer interface            : RS-232C
Scans to average              : 1
Surface PAR voltage added     : No
NMEA position data added      : No
Scan time added               : Yes

1) Frequency 0, Temperature
   Serial number : 4211
   Calibrated on : 08-Nov-07
   G             : 4.38706278e-003
   H             : 6.46536683e-004
   I             : 2.23272060e-005
   J             : 1.77872902e-006
   F0            : 1000.000
   Slope         : 1.00000000
   Offset        : 0.0000

2) Frequency 1, Conductivity
   Serial number : 2887
   Calibrated on : 18-Oct-07
   G             : -1.00439325e+001
   H             : 1.36330343e+000
   I             : -2.56289727e-003
   J             : 2.54823008e-004
   CTcor         : 3.2500e-006
   CPcor         : -9.57000000e-008
   Slope         : 1.00000000
   Offset        : 0.00000

3) Frequency 2, Pressure, Digiquartz with TC
   Serial number : 209
   Calibrated on : 09-Jul-07
   C1            : -3.920451e+004
   C2            : 6.234560e-001
   C3            : 1.350570e-002
   D1            : 3.894300e-002
   D2            : 0.000000e+000
   T1            : 3.046303e+001
   T2            : -9.018862e-005
   T3            : 4.528890e-006
   T4            : 3.309590e-009
   T5            : 0.000000e+000
   Slope         : 0.99985000
   Offset        : 1.00090
   AD590M        : 1.144000e-002
   AD590B        : -8.805040e+000

4) Frequency 3, Temperature, 2
   Serial number : 1455
   Calibrated on : 13-Nov-07
   G             : 4.84617647e-003
   H             : 6.77841857e-004
   I             : 2.60561588e-005
   J             : 2.02936086e-006
   F0            : 1000.000
   Slope         : 1.00000000
   Offset        : 0.0000

5) Frequency 4, Conductivity, 2
   Serial number : 2882
   Calibrated on : 18-Oct-07
   G             : -1.02006582e+001
   H             : 1.39961765e+000
   I             : 7.01158866e-004
   J             : 2.20787100e-005
   CTcor         : 3.2500e-006
   CPcor         : -9.57000000e-008
   Slope         : 1.00000000
   Offset        : 0.00000

6) A/D voltage 0, Oxygen, SBE 43
   Serial number : 315
   Calibrated on : 16-Oct-07p
   Equation      : Owens-Millard
   Coefficients for Owens-Millard: 
   Soc           : 3.6150e-001
   Boc           : 0.0000
   Offset        : -0.5838
   Tcor          : -0.0001
   Pcor          : 1.35e-004
   Tau           : 0.0
   Coefficients for Murphy-Larson: 
   Soc           : 0.00000e+000
   Offset        : 0.00000e+000
   A             : 0.00000e+000
   B             : 0.00000e+000
   C             : 0.00000e+000
   E             : 0.00000e+000
   Tau           : 2.00000e+000

7) A/D voltage 1, Free

8) A/D voltage 2, Free

9) A/D voltage 3, User Polynomial
   Serial number : 8756
   Calibrated on : 
   Sensor name   : Metrox
   A0            : 12.00000000
   A1            : 445.60000000
   A2            : 0.00000000
   A3            : 0.00000000

10) A/D voltage 4, Free

11) A/D voltage 5, Free

12) A/D voltage 6, Free

13) A/D voltage 7, Free
---------------------------------------------
Pump Control
   This setting is only applicable to a custom build of the SBE 9plus.
   Enable pump on / pump off commands: NO
---------------------------------------------
Data Acquisition:
   Archive data:                    YES
   Delay archiving:                 NO
   Data archive:                    C:P18_07data
h21731.hex
   Timeout (seconds) at startup:    20
   Timeout (seconds) between scans: 20
---------------------------------------------
Instrument port configuration:
   Port      = COM1
   Baud rate = 19200
   Parity    = N
   Data bits = 8
   Stop bits = 1
---------------------------------------------
Water Sampler Data:
   Water Sampler Type:   SBE Carousel
   Number of bottles:    24
   Port:                 COM2
   Enable remote firing: NO
   Firing sequence:      Sequential
---------------------------------------------
Header information:
   Header Choice = Prompt for Header Information
      prompt 0 = Cruise:  CLIVAR P18 2007
      prompt 1 = Ship:  NOAA RONALD H. BROWN
      prompt 2 = Station/Cast Number:
      prompt 3 = Nominal Latitude:
      prompt 4 = Nominal Longitude:
---------------------------------------------
TCP/IP - port numbers:
   Data acquisition:
      Data port:            49163
      Status port:          49165
      Command port:         49164
   Remote bottle firing:
      Command port:         49167
      Status port:          49168
   Remote data publishing: 
      Converted data port:  49161
      Raw data port:        49160
---------------------------------------------
Miscellaneous data for calculations
   Depth and Average Sound Velocity
      Latitude when NMEA is not available: 0.00000000
   Average Sound Velocity
      Minimum pressure [db]:               20.00000000
      Minimum salinity [psu]:              20.00000000
      Pressure window size [db]:           20.00000000
      Time window size [s]:                60.00000000
   Descent and Acceleration
      Window size [s]:                     2.00000000
   Plume Anomaly
      Theta-B:                             0.00000000
      Salinity-B                           0.00000000
      Theta-Z / Salinity-Z                 0.00000000
      Reference pressure [db]              0.00000000
   Oxygen
      Window size [s]:                     2.00000000
   Potential Temperature Anomaly
      A0:                                  0.00000000
      A1:                                  0.00000000
      A1 Multiplier:                       Salinity
---------------------------------------------
Serial Data Output:
   Output data to serial port: NO
---------------------------------------------
Mark Variables:
   Variables:
   Digits   Variable Name [units]
   ------   ---------------------
   0        Scan Count
   4        Pressure, Digiquartz [db]
   5        Temperature [ITS-90, deg C]
   5        Salinity [PSU]
   4        Oxygen, SBE 43 [umol/Kg]
   5        Density [sigma-theta, Kg/m^3]
---------------------------------------------
Shared File Output:
   Output data to shared file: NO
---------------------------------------------
TCP/IP Output:
   Raw data:
      Output raw data to socket:              NO
      XML wrapper and settings:               NO
      Seconds between raw data updates:       0.00000000
   Converted data:
      Output converted data to socket:        NO
      XML format:                             NO
---------------------------------------------
SBE 11plus Deck Unit Alarms
   Enable minimum pressure alarm:    NO
   Enable maximum pressure alarm:    NO
   Enable altimeter alarm:           NO
---------------------------------------------
SBE 14 Remote Display
   Enable SBE 14 Remote Display:     NO
---------------------------------------------
Options:
   Prompt to save program setup changes: YES
   Automatically save program setup changes on exit: NO
   Confirm instrument configuration change: YES
   Confirm display setup changes: YES
   Confirm output file overwrite: YES
   Check scan length: YES
   Compare serial numbers: YES
   Maximized plot may cover Seasave: NO