Instrument: CTD Sea-Bird 911

Dataset: 
Instrument: 
Instrument Short Name: CTD SBE 911
(http://vocab.nerc.ac.uk/collection/L22/current/TOOL0035/)
Instrument Description:

The Sea-Bird SBE 911 is a type of CTD instrument package. The SBE 911 includes the SBE 9 Underwater Unit and the SBE 11 Deck Unit (for real-time readout using conductive wire) for deployment from a vessel. The combination of the SBE 9 and SBE 11 is called a SBE 911. The SBE 9 uses Sea-Bird's standard modular temperature and conductivity sensors (SBE 3 and SBE 4). The SBE 9 CTD can be configured with auxiliary sensors to measure other parameters including dissolved oxygen, pH, turbidity, fluorescence, light (PAR), light transmission, etc.). More information from Sea-Bird Electronics.

PI supplied instrument name: CTD Sea-Bird 911
Dataset-specific description

Data was acquired using SeaBird SeaSave for SBE 911 software (v7)


Sample Header File

* Sea-Bird SBE 9 Data File:

* FileName = C:/CTD/balch-12-2008/06101.hex

* Software Version Seasave V 7.18b

* Temperature SN = 4307

* Conductivity SN = 2593

* Number of Bytes Per Scan = 44

* Number of Voltage Words = 5

* Number of Scans Averaged by the Deck Unit = 1

* Append System Time to Every Scan

* System UpLoad Time = Dec 17 2008 12:46:09

* NMEA Latitude = 48 45.00 S

* NMEA Longitude = 059 47.47 W

* NMEA UTC (Time) = Dec 17 2008 12:46:10

* Store Lat/Lon Data = Append to Every Scan

** Station/Cast: 06101

** LAT: 48 45.000S

** LON: 59 47.475W

** Bottom Depth: 466

** Cast Type: DN

** Event No: 200812171246

# units = specified

# bad_flag = -9.990e-29

# sensor 0 = Frequency 0 temperature, primary, 4307, 11-Sep-08

# sensor 1 = Frequency 1 conductivity, primary, 2593, 03-Sep-08 , cpcor = -9.5700e-08

# sensor 2 = Frequency 2 pressure, 0831, 17-Sep-08

# sensor 3 = Frequency 3 temperature, secondary, 2495, 11-Sep-08

# sensor 4 = Frequency 4 conductivity, secondary, 2766, 09-Sep-08 , cpcor = -9.5700e-08

# sensor 5 = Extrnl Volt 0 Fluorometer, Seapoint, primary

# sensor 6 = Extrnl Volt 1 irradiance (PAR), primary, 4644, 11 Feb. 2004

# sensor 7 = Extrnl Volt 2 transmissometer, primary, CST-1115-DR, 01 May 2008

# sensor 8 = Extrnl Volt 4 altimeter

# sensor 9 = Extrnl Volt 6 Oxygen, SBE, primary, 1129, 30-May-08p

# sensor 10 = Extrnl Volt 9 surface irradiance (SPAR), degrees = 0.0

# datcnv_date = Dec 17 2008 23:17:36, 7.18b

# datcnv_in = w:/06101.hex w:/06101.CON

# datcnv_skipover = 4800

# datcnv_ox_hysteresis_correction = yes

# datcnv_ox_tau_correction = yes

# wildedit_date = Dec 17 2008 23:17:41, 7.18b

# wildedit_in = w:/06101.cnv

# wildedit_pass1_nstd = 2.0

# wildedit_pass2_nstd = 10.0

# wildedit_pass2_mindelta = 0.000e+000

# wildedit_npoint = 500

# wildedit_vars = prDM t090C c0mS/cm t190C c1mS/cm sbeox0ML/L sbeox0Mm/Kg sbeox0PS flSP xmiss bat par spar cpar altM

# wildedit_excl_bad_scans = yes

# filter_date = Dec 17 2008 23:17:46, 7.18b

# filter_in = w:/06101.cnv

# filter_low_pass_tc_A = 0.030

# filter_low_pass_tc_B = 0.150

# filter_low_pass_A_vars =

# filter_low_pass_B_vars = prDM

# alignctd_date = Dec 17 2008 23:17:54, 7.18b

# alignctd_in = w:/06101.cnv

# alignctd_adv = sbeox0ML/L 7.000, sbeox0Mm/Kg 7.000, sbeox0PS 7.000

# celltm_date = Dec 17 2008 23:17:59, 7.18b

# celltm_in = w:/06101.cnv

# celltm_alpha = 0.0300, 0.0300

# celltm_tau = 7.0000, 7.0000

# celltm_temp_sensor_use_for_cond = primary, secondary

# loopedit_date = Dec 17 2008 23:18:03, 7.18b

# loopedit_in = w:/06101.cnv

# loopedit_minVelocity = 0.050

# loopedit_surfaceSoak: do not remove

# loopedit_excl_bad_scans = yes

# Derive_date = Dec 17 2008 23:18:10, 7.18b

# Derive_in = w:/06101.cnv w:/06101.CON# file_type = ascii

*END*

This mstar file created from sbe file 06101.cnv at 2008-12-18 11:22:36

condcal = 1.000186 * cond

cond2cal = 1.000098 * cond2

oxygen calibration = 0.058575 + 1.048204 * sbeoxy