Instrument: Flow injection lumogallion system with fluorometer

Instrument Short Name: FIA with Fluorometer
(http://vocab.nerc.ac.uk/collection/L22/current/TOOL0513/)
Instrument Description:

An analytical system used to determine concentrations of chemical species in a sample based on the fluorescence from the reaction between lumogallion and the species of interest. The system typically comprises individual components typically including pumps, injection and autosampler valves, preconcentration columns and a fluorometer. The system is normally uniquely assembled for each analysis.
(From SeaDataNet)

PI supplied instrument name: Flow injection lumogallion system with fluorometer
Dataset-specific description

Dissolved Al was determined using a Flow Injection Analysis scheme with fluorometeric detection. The scheme produces a complex between lumogallion and dissolved Al. Major components of the scheme include a Rabbit peristaltic pump, a Dynamax FL-1 fluorometer, a Rainin A/D board, and a Macintosh G3 computer running Rainin MacIntegrator v 1.4.3 to log and reduce data.