/BCO/SO-GasEx/RB-08-02/Underway_O2_Ar_Log ---- Level 0

      Directory  DocumentationPlotting and Other Operations...

 - At level 0 -       Flat list

#  Version: 14 October 2013
#  PIs: Hamme, Bender, Cassar
#  Final Underway O2/Ar - Sampling Log
# 
=========================
Date_Time_GMT                        Event_or_Comment                                                                                                                                                                                                                                                                     
-------------------------
1-Mar-08 13:14                       Began collecting data.  Sampling air only.  Switching between capilaries every 10 minutes                                                                                                                                                                                            
2-Mar-08 03:05                       System off                                                                                                                                                                                                                                                                           
2-Mar-08 11:55                       System on.  Still sampling air.                                                                                                                                                                                                                                                      
2-Mar-08 13:35                       Continuous seawater flow on.  Flowing through EIMS system.                                                                                                                                                                                                                           
2-Mar-08 14:50                       Oxygen optode placed in seawater.                                                                                                                                                                                                                                                    
2-Mar-08 17:10                       Changed valve switching to repeating pattern of 10 min air; 20 min equilibrator.                                                                                                                                                                                                     
2-Mar-08 20:33 to 2-Mar-08 20:55     Seawater flow off and optode out of water while tubing upstream of equilibrator was shortened.                                                                                                                                                                                       
2-Mar-08 21:30                       Adjusted optode position deeper in flowing water.  Temperature measured at optode decreased.                                                                                                                                                                                         
3-Mar-08 12:00                       Changed valve switching to repeating pattern of 10 min air; 50 min equilibrator.                                                                                                                                                                                                     
3-Mar-08 15:15 to 3-Mar-08 15:35     Seawater flow off and optode out of water while insulating tubing upstream of equilibrator.                                                                                                                                                                                          
4-Mar-08 00:36 to 4-Mar-08 21:18     Lost data from quadrupole mass spec during this time due to file use interference.                                                                                                                                                                                                   
4-Mar-08 10:00                       Changed valve switching to repeating pattern of 10 min air; 1 hr 50 min equilibrator.                                                                                                                                                                                                
4-Mar-08 21:35 to 4-Mar-08 21:50     Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
5-Mar-08 16:49 to 5-Mar-08 17:00     Adjusting large flow of water.  Optode out of water.  Flow to equilibrator off.                                                                                                                                                                                                      
5-Mar-08 20:02 to 5-Mar-08 20:09     Seawater flow off while adding valves to system.  Optode out of water.  Flow to equilibrator off.                                                                                                                                                                                    
7-Mar-08 22:35                       Underway system off for tracer injection.  Forced capillary to air sampling.  Optode remained in water; but no flow past it.                                                                                                                                                         
7-Mar-08 23:45                       Breathed on air inlet.  CO2 increased.  O2 decreased.                                                                                                                                                                                                                                
8-Mar-08 06:30 to 8-Mar-08 15:00     Quadrupole oxygen current flipping back and forth between two slightly different values as decreasing current brings it over AtoD converter threshold.                                                                                                                               
8-Mar-08 15:30                       Performed mass calibration on quadrupole.                                                                                                                                                                                                                                            
8-Mar-08 16:36                       Cleaning / changing filters.                                                                                                                                                                                                                                                         
8-Mar-08 17:40                       Filled graduated cylinder and bucket with fresh water.  Small pump on.  Recycling same water.                                                                                                                                                                                        
8-Mar-08 18:24                       Moved pump inlet to bucket to provide larger reservoir.                                                                                                                                                                                                                              
9-Mar-08 00:09                       Changed back to continuous seawater flow.                                                                                                                                                                                                                                            
9-Mar-08 02:10 to 9-Mar-08 02:20     Repeat of oxygen AtoD converter threshold problem.                                                                                                                                                                                                                                   
10-Mar-08 12:10 to 10-Mar-08 18:00   3 emission current faults in quadrupole.  Filament turned itself off.                                                                                                                                                                                                                
11-Mar-08 11:24 to 11-Mar-08 14:28   Quadrupole data loss                                                                                                                                                                                                                                                                 
11-Mar-08 20:03 to 11-Mar-08 20:08   Cleaning out black filter. Small pump flow off.                                                                                                                                                                                                                                      
13-Mar-08 01:55                      Emission error turned off filament in quadrupole again.  System problems being diagnosed.  Seawater flows are off.                                                                                                                                                                   
13-Mar-08 18:00                      Seawater flows back on.  Optode measuring again.  Quadrupole still not producing reliable data.                                                                                                                                                                                      
15-Mar-08 18:00                      Continued testing.  Optode removed from seawater flow.                                                                                                                                                                                                                               
13-Mar-08 01:55 to 16-Mar-00 00:00   Things changed as part of diagnosis testing of quadrupole:  Switched to filament #2.  Replaced all capillaries.  Opened and cleaned valco valve.  Switched to larger size ferrules in valco valve.  Replaced equilibrator cartride; which was probably the main reason for the fix.  
16-Mar-08 00:00                      Back to normal operation                                                                                                                                                                                                                                                             
17-Mar-08 00:13                      Moved quad housing temperature sensor to somewhat different position inside housing.                                                                                                                                                                                                 
17-Mar-08 15:30                      Lab air thermistor was pulled out of Omega AtoD converter.  Screwed back into terminals.                                                                                                                                                                                             
19-Mar-08 01:59 to 20-Mar-08 19:57   Periodic bursts of air bubbles in underway system.  Causes spikes in optode oxygen sensor and N2/Ar ratio.  No apparent effect on O2/Ar expect for really large slugs of air.                                                                                                        
21-Mar-08 04:04                      Seawater flow off for tracer injection.  Quadrupole sampling air only.  All thermistors gathered together in air to test for offsets.                                                                                                                                                
21-Mar-08 18:00                      Changed equilibrator cartridge. Back to normal operation                                                                                                                                                                                                                             
21-Mar-08 20:30                      Emission error turned off filament in quadrupole again.  Seawater flows are off.  Equilibrator cartridge full of water and drips steadily with pump on; must be faulty.                                                                                                              
21-Mar-08 20:30 to 22-Mar-08 16:00   Switched to new equilibrator cartridge.  Continued problems. Discovered capillary was extending far into cartridge and likely touching membrane.  Reterminated capilary and seated near swagelok fitting.                                                                            
22-Mar-08 16:00                      System returned to normal operation.                                                                                                                                                                                                                                                 
23-Mar-08 11:40 to 26-Mar-08 00:37   Periodic bursts of air bubbles in underway system.  Causes spikes in optode oxygen sensor and N2/Ar ratio.  No apparent effect on O2/Ar expect for really large slugs of air.                                                                                                        
23-Mar-08 16:50 to 23-Mar-08 18:00   Signal may be compromised by intake of ship sewage.                                                                                                                                                                                                                                  
27-Mar-08 15:09 to 27-Mar-08 15:30   Cleaning / changing filters.  Optode out of water.  Flow off.  Apparent step change in O2/Ar at this time.  Likely too long since filters were changed and system was experiencing internal respiration.                                                                             
28-Mar-08 18:02 to 28-Mar-08 18:10   Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
29-Mar-08 11:13 to 31-Mar-08 09:10   Periodic bursts of air bubbles in underway system.  Causes spikes in optode oxygen sensor and N2/Ar ratio.  No apparent effect on O2/Ar expect for really large slugs of air.                                                                                                        
29-Mar-08 20:00 to 29-Mar-08 20:06   Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
31-Mar-08 00:02 to 31-Mar-08 00:20   Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
1-Apr-08 00:02 to 1-Apr-08 00:20     Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
1-Apr-08 05:43 to 1-Apr-08 06:20     Continuous seawater system pump failure.                                                                                                                                                                                                                                             
3-Apr-08 00:02 to 3-Apr-08 00:15     Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
3-Apr-08 04:48 to                    Periodic bursts of air bubbles in underway system.  Causes spikes in optode oxygen sensor and N2/Ar ratio.  No apparent effect on O2/Ar expect for really large slugs of air.                                                                                                        
5-Apr-08 00:03 to 5-Apr-08 00:15     Cleaning / changing filters.  Optode out of water.  Flow off.                                                                                                                                                                                                                        
5-Apr-08 16:04 to 5-Apr-08 16:30     Continuous seawater system pump failure.