PSA file: C:\Data\ctd\193-03A.psa Date: 05/16/2008 Instrument configuration file: C:\Data\ctd\19303090.CON Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : Yes NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 2774 Calibrated on : 14/feb/2008 G : 4.32553808e-003 H : 6.46845042e-004 I : 2.40832616e-005 J : 2.45256750e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 1474 Calibrated on : 14/feb/2008 G : -4.20159424e+000 H : 5.37564619e-001 I : -1.87354571e-004 J : 3.89029332e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 63505 SBE090462 Calibrated on : 12/23/2002 C1 : -4.872453e+004 C2 : 2.143123e-002 C3 : 1.347220e-002 D1 : 3.959500e-002 D2 : 0.000000e+000 T1 : 2.994567e+001 T2 : -2.488396e-004 T3 : 3.985300e-006 T4 : 7.998620e-010 T5 : 0.000000e+000 Slope : 0.99992000 Offset : -1.29630 AD590M : 1.282050e-002 AD590B : -9.111539e+000 4) Frequency 3, Temperature, 2 Serial number : 2900 Calibrated on : 14/feb/08 G : 4.32909519e-003 H : 6.44031984e-004 I : 2.32248662e-005 J : 2.23696849e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 1859 Calibrated on : 04/march/2008 G : -4.11374373e+000 H : 5.20688734e-001 I : -8.43469509e-004 J : 6.87043888e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Oxygen, SBE 43 Serial number : 0113 Calibrated on : 13/march/08 Equation : Owens-Millard Coefficients for Owens-Millard: Soc : 3.1150e-001 Boc : 0.0000 Offset : -0.5783 Tcor : 0.0022 Pcor : 1.35e-004 Tau : 0.0 Coefficients for Murphy-Larson: Soc : 0.00000e+000 Offset : 0.00000e+000 A : 0.00000e+000 B : 0.00000e+000 C : 0.00000e+000 E : 0.00000e+000 Tau : 2.00000e+000 7) A/D voltage 1, Fluorometer, Wetlab CDOM Serial number : WSCD-1094P Calibrated on : 10/sept/2007 Vblank : 0.060 Scale factor : 51.000 8) A/D voltage 2, Free 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : 70166 Calibrated on : 2/26/08 M : 1.00000000 B : 0.00000000 Calibration constant : 12406900000.00000000 Multiplier : 1.00000000 Offset : 0.00000000 10) A/D voltage 4, User Polynomial Serial number : FLNTU-873 Calibrated on : 12/10/2007 Sensor name : Scattering Meter A0 : 0.00000000 A1 : 0.00000000 A2 : 0.00000000 A3 : 0.00000000 11) A/D voltage 5, User Polynomial, 2 Serial number : FL3-302 Calibrated on : 12/10/07 Sensor name : ECO F FLNTU A0 : 0.00000000 A1 : 0.00000000 A2 : 0.00000000 A3 : 0.00000000 12) A/D voltage 6, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : CST-1090PR Calibrated on : March 17, 2008 M : 22.1991 B : -1.2469 Path length : 0.250 13) A/D voltage 7, Free 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 6294 Calibrated on : 03/13/2008 Conversion factor : 1827.00000000 Ratio multiplier : 1.00000000 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: C:\Data\ctd\19303091.hex Timeout (seconds) at startup: 30 Timeout (seconds) between scans: 10 --------------------------------------------- Instrument port configuration: Port = COM3 Baud rate = 19200 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: SBE Carousel Number of bottles: 24 Port: COM4 Enable remote firing: NO Firing sequence: Sequential Tone for bottle fire confirmation uses PC sound card. --------------------------------------------- Header information: Header Choice = Prompt for Header Information prompt 0 = Ship: R/V Knorr prompt 1 = Cruise: 193-03 prompt 2 = Station: --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth and Average Sound Velocity Latitude when NMEA is not available: 0.00000000 Average Sound Velocity Minimum pressure [db]: 20.00000000 Minimum salinity [psu]: 20.00000000 Pressure window size [db]: 20.00000000 Time window size [s]: 60.00000000 Descent and Acceleration Window size [s]: 2.00000000 Plume Anomaly Theta-B: 0.00000000 Salinity-B 0.00000000 Theta-Z / Salinity-Z 0.00000000 Reference pressure [db] 0.00000000 Oxygen Window size [s]: 2.00000000 Potential Temperature Anomaly A0: 0.00000000 A1: 0.00000000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: NO --------------------------------------------- Mark Variables: Variables: Digits Variable Name [units] ------ --------------------- 0 Scan Count 4 Depth [salt water, m] 7 Conductivity [S/m] 5 Salinity [PSU] --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: NO XML wrapper and settings: NO Seconds between raw data updates: 0.00000000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC internal speakers. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO