Custom-built Winkler dissolved oxygen titrator: Documentation and code developed and used for oxygen titrations (Nicholson et al., 2023: https://doi.org/10.5281/zenodo.8048208)
Custom-built Winkler dissolved oxygen titrator: Documentation and code developed and used for oxygen titrations for this project (Nicholson et al., 2023: https://doi.org/10.5281/zenodo.8048208)
All samples were titrated within one week of collection at the Boston College Marine Biogeochemistry Lab using a custom-built Winkler titrator with automated potentiometric end point detection (control software available here: Nicholson et al., 2023).
The oxygen probe measurements are regularly (at least monthly) compared and calibrated (if necessary) against the high-precision Winkler titrations using a (Parsons et al., 1984; Hansen, 1999) and a Brinkmann Metrohm titrator